Dimension (Excl. ancillaries) |
150 x 88 x 50mm |
Weight |
485 grams |
Power supply |
1 x 3.6V AA Lithium battery |
Case material |
ABS & PC |
Memory capacity |
50,000 readings per channel (unit will wrap when full) |
Clock accuracy (logging) |
20 ppm at 25°C |
IP Rating |
IP30 |
Instrument operating range |
-20°C to +65°C in a non-condensing RH environment |
Storage temperature |
-40°C to +60°C |
Frequency options |
A range of frequencies are available between 433-458MHz. Country specific regulations apply |
Radio power |
10mW |
Radio range |
3km over open ground |
Battery life |
Up to 2 years (dependent on conditions of use and instrument settings) |
Software required |
W900 – Standard EMS Software Package
W906 – Validated EMS Software Package |
Hardware required |
CR2 / CR3 – Controller
SR2 – Smart Receiver
REP – Repeater |
Accessories |
Wall mounting bracket
AA Lithium battery |
Sensor options (supplied with unit) |
J140 EE07-02 RH/T probe
Temperature Probe: PT1000 (Tolerance class A)
Temperature operating range:
-40°C to +80°C (restricted by
instrument operating range)
Cable operating range: -40°C to +80°C
Temperature accuracy: +/- 0.1°C at 20°C
Long term drift: +/- 0.05°C per year
Humidity Probe: Capacitive Polymer
Recommended range: 0-100% RH non condensing
Accuracy +/- 2% (0-90%)
+/- 3% (90-100%)
Temperature dependance: <(0.025+0.0003xRH)[%RH/°C] |
Solar and rain shield |
T351-G18
OPERATING ENVIRONMENT
Operating/Storage
temperature: -40°C to +80°C
Maximum wind speed: 75 m/s (168mph)
SHIELDING PERFORMANCE
Shielding efficiency
(temperature rise/error due
to incident solar radiation):
2.3°C/KW/m² at 1m/s typical
MECHANICS
Diameter: 76mm
Shield materials: Nylon/polyamide, polycarbonate,
ABS, aluminium, stainless-steel
Overall height: 118mm
Typical weight: 110g |
Price |
 |